• 4-20mA PT100 Temperature Transmitter
  • 4-20mA PT100 Temperature Transmitter
  • 4-20mA PT100 Temperature Transmitter
  • 4-20mA PT100 Temperature Transmitter
  • 4-20mA PT100 Temperature Transmitter
  • 4-20mA PT100 Temperature Transmitter

4-20mA PT100 Temperature Transmitter

Type: Wall-Mounted Current Type
Measuring Medium: Temperature
Display Type: None
Site Display: None
Accuracy: 0.5
IP Rating: IP65
Samples:
US$ 15/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Diamond Member Since 2021

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Certification
CE
Customized
Customized
Transport Package
Carton Box
Specification
1kg
Trademark
XYT
Origin
China
HS Code
9026
Production Capacity
2000/Month

Product Description

3 Wire Rtd 4-20mA PT100 Temperature Transmitter
 
4-20mA PT100 Temperature Transmitter4-20mA PT100 Temperature Transmitter
4-20mA PT100 Temperature Transmitter
Product Introduction

The intelligent temperature transmitter (round card) is used for the input of thermal resistance (RTD) and thermocouple (TC) signals and the output of 2-wire 4-20ma analog system. It is installed inside the sensor (Form B).
1.Dimension
4-20mA PT100 Temperature Transmitter
2. Installation Diagram
4-20mA PT100 Temperature Transmitter
3.Configuration diagram
4-20mA PT100 Temperature Transmitter
4.Wiring diagram4-20mA PT100 Temperature Transmitter
 
5. Specification
Enter
input signal Thermal resistance (RTD), thermocouple (TC)
Cold junction compensation range -20~60ºC
Compensation accuracy ±1ºC
Output
output signal 4~20mA  
Load Resistance RL≤(Ue-12)/0.021
Upper and Lower limit overflow alarm output current IH=21mA,IL=3.8mA
Power supply
Supply voltage DC12~40V
Other parameters
Temperature drift 0.02%FS/ºC
Response time 1s reaches 90% of final value
Working temperature -40~80ºC
Stored temperature -40~100ºC
Shock resistance 4g/2~150Hz
installation angle Unlimited
Installation area b-type top box mounting
Electromagnetic Meet the requirements of gb / t18268 industrial equipment
  1. Input type and transmission accuracy
    Model Types Measuring Minimum Measuring Conversion

    RTD
    Pt100 -200.0~850.0ºC 10ºC ≤100ºC 0.2%
    >100ºC 0.1%
    Cu50 -50.0~150.0ºC 10ºC ≤100ºC 0.2%
    >100ºC 0.1%







    TC
    B 400~1800ºC 500ºC ≤300ºC 0.2%
    >300ºC 0.1%
    E -100~1000ºC 50ºC ≤300ºC 0.2%
    >300ºC 0.1%
    J -100~1200ºC 50ºC ≤300ºC 0.2%
    >300ºC 0.1%
    K -180~1372ºC 50ºC ≤300ºC 0.2%
    >300ºC 0.1%
    N -180~1300ºC 50ºC ≤500ºC 0.2%
    >500ºC 0.1%
    R -50~1768ºC 500ºC ≤500ºC 0.2%
    >500ºC 0.1%
    S -50~1768ºC 500ºC ≤500ºC 0.2%
    >500ºC 0.1%
    T -200~400ºC 50ºC ≤500ºC 0.2%
    >500ºC 0.1%
    Wre3-25 0~2315ºC 500ºC ≤500ºC 0.2%
    >500ºC 0.1%
    Wre5-26 0~2310ºC 500ºC ≤500ºC 0.2%
    >500ºC 0.1%
Note:
1. The above accuracy data is obtained by testing under the environment temperature of 20 ºC
± 2 ºC.
2. The output accuracy "%" is relative to the set range.
3.The cold junction compensation error needs to be added when measuring the thermocouple.